This listing is for a PALL Palltronic Compact Star.
The unit is in full working condition and is ready for immidate realese.
Supplied with 2 replacment Epsom Ribbon Cartridges and software disk.
An automated integrity test, whether forward flow or pressurehold, is internationally acknowledged as a truly objective and non-destructive method for determining if a filter meets microbial re-moval specifications and is correctly installed. Test results providea clear and accurate indication of a filter’s integrity prior to its use.
Integrity testing with the Palltronic Compact Star device is reliable,cost-effective, and operator-friendly. The unit is designed to per-form a Pressure Hold Test, a nondestructive method that measurespressure decay over a period of time. The test involves automati-cally pressurizing the filter housing at a predetermined level, allow-ing the pressure to stabilize. The unit then takes the test measure-ment and depressurizes the assembly through the vent valve. Fullyautomated, the Palltronic Compact Star device eliminates the vari-ability of test results that can occur with manual systems. It pro-vides a simple, fast, and objective check of filter integrity
Test method (test pressure): Pressure Hold TestMeasuring range:50 mbar – 6000 mbar (0.7 psi – 87 psi)Pressure supply:max. 8000 mbar (87 psi)Min. 1000 mbar (14.5 psi) over test pressureThe pressure supply has to be clean and dry(