USER COMMAND PARAMETERS
• (Servo) control of newborn temperature: 35 to 37.5 °C in 0.1 ° increments
(regulates the infant's temperature whether the bed is in open or closed mode)
• Manual radiation power control: 0 to 100% in 5% increments
• Humidity Servo Control Range: 30 to 95% relative humidity in 5% increments System Performance
• Expected life: This device is expected to last at least seven years under normal use, provided it is used and maintained in accordance with the instructions given in the manuals provided. This life can be extended by regular maintenance and proper repairs, as long as replacement parts are available.
• Closed bed performance, incubator mode: Measurement accuracy at patient level: ±0°C between 30° and 42°C
• Oxygen servo control performance
• Ramp-up time: <10 minutes from % below setpoint in Whisper Quiet mode
• Recovery after opening a porthole: ≤ 5 minutes after closing the porthole up to 5% below the set point
• 5% accuracy Maximum inlet pressure of 620 kPa (90 psi) Humidifier performance