4.1 General description
Cutting with automatic HF voltage control (Auto Cut)
The ERBOTOM ICC is equipped with open and closed loop control systems
to control and regulate the parameters relevant to the cutting quality so that each
cutting quality is guaranteed to be reproducible and constant.
Soft coagulation:
Soft coagulation can be activated by a key or a foot pedal.
Forced coagulation:
Forced coagulation is advantageous if effective hemostasis is to be achieved with relatively small surface electrodes, such as TUR resection loops.
Spray coagulation:
The spray coagulation mode on the ICC is also suitable for argon gas coagulation.
Adjustable power limitation in the different coagulation modes
For ERBOTOM ICC units, the surgically relevant coagulation qualities, i.e. the coagulation effects Soft Coag., Forced Coag. Spray Coag. and Bipolar Coag., are delimited by definition from each other and selectable by pressing a button. Nevertheless, the intensity of the different effects can be changed by the power limitation.
Bipolar coagulation:
In this coagulation mode, the HF voltage is, similar to soft coagulation, automatically and constantly controlled, and its maximum value always remains below 200 Vp, so that the current density and thus the coagulation effect is, for the most part, independent of the effective contact surface between the coagulation electrode and the tissue, provided that the effective contact surface is not too large compared to the currently set power limitation.